Eric2019 发表于 2019-8-2 15:25:27

有修Medrad高压注射器的吗

求Medrad stellant D 高压注射资料,或高手指点一二。开关报SW311是什么原因???

tsect 发表于 2019-8-3 08:52:07

SW - Software Problem
Errors with SW typically indicate a problem with internal software. This may be a condition that
must be corrected, or simply an electrical transient that disrupted the system. Try the following:
• Remove and re-apply power to the unit. This triggers a self-test that will attempt to verify
proper unit operation. If the problem is corrected, there was probably a voltage transient in
the unit, now cleared, enabling system use.
• If the problem continues, contact MEDRAD to reload software.
• If the problem continues, replace the SBC card.
• If the problem continues, replace the Servo/CPU card.

Eric2019 发表于 2019-8-3 13:58:11

tsect 发表于 2019-8-3 08:52
SW - Software Problem
Errors with SW typically indicate a problem with internal software. This may...

你好,谢谢指点。开机10次就会有三四次报SW311,如果不报311就能正常使用。把这注射器个注射器换到另外一台主机上也一样报错。这跟软件有关吗???你说的reload software是对主机还是注射器?

tsect 发表于 2019-8-5 07:18:14

sorry,这个我是从某个手册上看的,没有具体操作说明,个人理解,reload software应该是对注射器。

tsect 发表于 2019-8-5 07:37:12

可以先查一下相关板卡的供电是否正常

tsect 发表于 2019-8-5 07:46:46

包括与CPU卡的通讯电缆

只要你好 发表于 2020-6-24 10:44:38

解决了吗