有修Medrad高压注射器的吗
求Medrad stellant D 高压注射资料,或高手指点一二。开关报SW311是什么原因??? SW - Software ProblemErrors with SW typically indicate a problem with internal software. This may be a condition that
must be corrected, or simply an electrical transient that disrupted the system. Try the following:
• Remove and re-apply power to the unit. This triggers a self-test that will attempt to verify
proper unit operation. If the problem is corrected, there was probably a voltage transient in
the unit, now cleared, enabling system use.
• If the problem continues, contact MEDRAD to reload software.
• If the problem continues, replace the SBC card.
• If the problem continues, replace the Servo/CPU card. tsect 发表于 2019-8-3 08:52
SW - Software Problem
Errors with SW typically indicate a problem with internal software. This may...
你好,谢谢指点。开机10次就会有三四次报SW311,如果不报311就能正常使用。把这注射器个注射器换到另外一台主机上也一样报错。这跟软件有关吗???你说的reload software是对主机还是注射器?
sorry,这个我是从某个手册上看的,没有具体操作说明,个人理解,reload software应该是对注射器。 可以先查一下相关板卡的供电是否正常 包括与CPU卡的通讯电缆
解决了吗