松下MLS高压蒸汽灭菌器
松下MLS-3751L-PC高压蒸汽灭菌器开机报警E5是什么意思?谁有说明书或者报警代码?谢谢E5
Bottom thermistor (TH3)
• Higher than 110'C in heating I,
• Check the thermistor (TH3).
Abnonnal tamp
keep-warm process.
• Check whether or not the exhaust port
• Higher than 120'C in heating 2,
is blocked with sterilization object.
exhaust process.
• Isn't the amount of heating water
• Higher than 140'C in heating 3, heating
reduced and isn't the chamber is
4, sterilization process.
heated wnhout water?
• Or, higher than set temp.+30·C
• Check for leakage of steam or water. tsect 发表于 2024-8-22 19:08
E5
Bottom thermistor (TH3)
十分感谢
学习了:victory::victory::victory::victory::victory: tsect(牛)