史赛克x8000

  [复制链接]
已绑定手机
发表于 2021-5-7 15:47:15 来自手机 | 显示全部楼层 |阅读模式
史赛克X8000出现e1 E3什么意思
发表于 2021-5-7 22:17:52 | 显示全部楼层
E-1  All conditions are met for the bulb toilluminate, yet it remains off.         Return the X8000 forrepair
E-2  All conditions are not met for the bulbto illuminate, yet it remains on.   Return the X8000 forrepair
E-3  The bulb has higher than expectedcurrent or voltage applied to it.    .   Return the X8000 forrepair
E-4  The bulb is kept off because the ballastfan is not working properly.      Return the X8000 forrepair
E-5  The bulb is kept off because the bulb fanis not working properly.         
Blinking500  The bulb has exceeded its recommended lifetime of 500 hours.     .Install a new Stryke
发表于 2021-5-7 23:07:21 | 显示全部楼层
E-1 error appears on the LCD display
1.  Make sure that the light cable is inserted, the bulb
door is closed with a bulb inside, and the lamp is
not lit.
2.  Check the voltages on TP22 and TP23.
•  If the voltage diference between TP22 and
TP23 is not 12V, replace the control board.
•  If the voltage diference between these two test
points is 12V, ensure the J3 cable is connected
properly to the ballast and that the individual
wires are not crossed.
-  Ensure that L3 and L4 ferrites, located
underneath the J3 connector, are
soldered properly to the control board.
-  If these ferrites are missing, broken, or
not soldered properly, repair or replace
the control board.
3.  If none of the above steps resolves the problem,
replace the ballast.
发表于 2021-5-7 23:07:45 | 显示全部楼层
E-3 error appears on the LCD display
1.  Probe TP13 with a multimeter.
•  If it reads 3.3 volts, and E-3 still appears on the
display, replace the control board.
•  If the voltage on TP13 is low, replace the
ballast.
发表于 2022-8-1 13:27:44 | 显示全部楼层
有没有中文的?
已绑定手机
发表于 2023-5-8 05:49:57 来自手机 | 显示全部楼层
tsect 发表于 2021-5-7 23:07
E-1 error appears on the LCD display
1.  Make sure that the light cable is inserted, the bulb
door i ...

我也出现过,主要还是光纤没接好。