星图源医疗 来自手机 显示全部楼层 发表于 2021-5-7 15:47:15 |阅读模式
史赛克X8000出现e1 E3什么意思
tsect 显示全部楼层 发表于 2021-5-7 22:17:52
E-1  All conditions are met for the bulb toilluminate, yet it remains off.         Return the X8000 forrepair
E-2  All conditions are not met for the bulbto illuminate, yet it remains on.   Return the X8000 forrepair
E-3  The bulb has higher than expectedcurrent or voltage applied to it.    .   Return the X8000 forrepair
E-4  The bulb is kept off because the ballastfan is not working properly.      Return the X8000 forrepair
E-5  The bulb is kept off because the bulb fanis not working properly.         
Blinking500  The bulb has exceeded its recommended lifetime of 500 hours.     .Install a new Stryke
tsect 显示全部楼层 发表于 2021-5-7 23:07:21
E-1 error appears on the LCD display
1.  Make sure that the light cable is inserted, the bulb
door is closed with a bulb inside, and the lamp is
not lit.
2.  Check the voltages on TP22 and TP23.
•  If the voltage diference between TP22 and
TP23 is not 12V, replace the control board.
•  If the voltage diference between these two test
points is 12V, ensure the J3 cable is connected
properly to the ballast and that the individual
wires are not crossed.
-  Ensure that L3 and L4 ferrites, located
underneath the J3 connector, are
soldered properly to the control board.
-  If these ferrites are missing, broken, or
not soldered properly, repair or replace
the control board.
3.  If none of the above steps resolves the problem,
replace the ballast.
tsect 显示全部楼层 发表于 2021-5-7 23:07:45
E-3 error appears on the LCD display
1.  Probe TP13 with a multimeter.
•  If it reads 3.3 volts, and E-3 still appears on the
display, replace the control board.
•  If the voltage on TP13 is low, replace the
ballast.
小马哥_gf228 显示全部楼层 发表于 2022-8-1 13:27:44
有没有中文的?
已绑定手机
Tyx 来自手机 显示全部楼层 发表于 2023-5-8 05:49:57
tsect 发表于 2021-5-7 23:07
E-1 error appears on the LCD display
1.  Make sure that the light cable is inserted, the bulb
door i ...

我也出现过,主要还是光纤没接好。