crdrct 发表于 2018-5-5 11:54:36

求锐柯5850相机维修手册

求锐柯5850相机维修手册

株洲刘工 发表于 2018-5-22 11:37:15

Description
This error occurs when the LOCAL PANEL does not receive communications
from the Machine Control System (MCS). The LOCAL PANEL waits for 50
seconds, then generates a fake button press in an attempt to provoke
communications from the MCS. After an additional 10 seconds, if the LOCAL
PANEL has not received MCS communications, this error is displayed on the
LOCAL PANEL User Interface. The error clears when the LOCAL PANEL
detects MCS communications.
System Impact
Unknown. If the MCS is down, the printer will not print films. If the problem is
limited to communications, the printer may be able to print films.
User Action
1. Shut down the IMAGER, then start the IMAGER.
2. If the error persists, call for service.

tsect 发表于 2018-5-7 07:19:57

参考一下本论坛页面:http://bbs.bmeol.com/thread-16469-1-1.html

GJG 发表于 2018-5-22 09:36:25

报警36935是什么意思

GJG 发表于 2018-5-22 09:47:27

tsect 发表于 2018-5-7 07:19
参考一下本论坛页面:http://bbs.bmeol.com/thread-16469-1-1.html

没有金币下不了

mri_xn 发表于 2018-5-22 22:34:29

非常感谢楼主