求助金宝报5.05 002
本院一台200s机器,自检过后一上病人就报除气压低5.05 002,观察除气压为-440至-460左右,设置为-500,除气泵占空比由62%慢慢升到75%左右后变为零,又循环反复,换过传感器,除气泵,moto阀,管道硅胶头,fm cpu 板均无果,请问高手有没遇到该故障,求指点。 本帖最后由 luobju 于 2018-7-27 19:46 编辑FCN 5.05 002 TOO HIGH DEGASS PRESSURE
General description of conditions for occurring:
The measured degassing pressure is greater than the set degassing
pressure for 15 minutes, and it is also greater than -500 mmHg for
15 minutes.
Technical description of conditions for occurring:
(FRI(FI_DEGASSPR) > ORI(OI_DEGASS$PR$SET) + 200) and
FRI((FI_DEGASSPR) > -5000)
for 150 continuous seconds and after 300s, 450s, 600s, 750s and 900s.
1 unit of FRI(FI_DEGASSPR) = 0.1mmHg.
Some possible explanations why the error has occurred.
* The degassing motor is broken.
* The degassing pump is not ok.
* The degassing pressure transducer is broken.
* The degassing pressure amplifier is broken.
* The A/D system is broken.
最后五条是故障原因。
页: