tfq6909 发表于 2012-4-19 17:54:36

求助,,,

我们卫生站的KODAKDV5800突然红灯亮了,报警号是36935,说明书里没有3开头的,请各位师傅帮我一下,谢谢了。

tfq6909 发表于 2012-4-19 19:19:08


Condition Code: 36-935
Description
This error occurs when the LOCAL PANEL does not receiv e communications
from the Machine Control System (MCS). The LOCAL PANEL waits for 50
seconds, then generates a fake butto n press in an attempt to provoke
communications from the MCS. After an additiona l 10 seconds, if the LOCAL
PANEL has not received MCS communications, this error is displayed on the
LOCAL PANEL User Interface. The error clears when the LOCAL PANEL
detects MCS communications.
System Impact
Unknown. If the MCS is down, the printer will not print films. If the problem is
limited to communications, the printer may be able to print films.
User Action
1. Shut down the IMAGER, then start the IMAGER.
2. If the error persists, call for service.
FE Action
Web Portal Message Log Message DICOM Message Severity
Local Panel: No
Communications from MCS
Local Panel not receiving
communications from MCS.
NORMAL 2-Warning

魂斗罗 发表于 2012-4-26 10:12:12

:L怎么全英文,没注释

Mat 发表于 2012-4-26 15:03:56

什么情况啊

xudaf0206 发表于 2012-4-29 16:52:48

什么情况啊
页: [1]
查看完整版本: 求助,,,