kodak DV5800
我们卫生站的KODAKDV5800突然红灯亮了,报警号是32936,说明书里没有3开头的,请各位师傅帮我一下,谢谢了。 哪位好心人帮帮我 拆了,拿无水酒精清洗一下激光头 另外,两个片盒被锁住,不能抽出来。:L 先感谢你。我试试。:) 你好,我试了,没用。错误代码是36935
Condition Code: 36-935
Description
This error occurs when the LOCAL PANEL does not receiv e communications
from the Machine Control System (MCS). The LOCAL PANEL waits for 50
seconds, then generates a fake butto n press in an attempt to provoke
communications from the MCS. After an additiona l 10 seconds, if the LOCAL
PANEL has not received MCS communications, this error is displayed on the
LOCAL PANEL User Interface. The error clears when the LOCAL PANEL
detects MCS communications.
System Impact
Unknown. If the MCS is down, the printer will not print films. If the problem is
limited to communications, the printer may be able to print films.
User Action
1. Shut down the IMAGER, then start the IMAGER.
2. If the error persists, call for service.
FE Action
Web Portal Message Log Message DICOM Message Severity
Local Panel: No
Communications from MCS
Local Panel not receiving
communications from MCS.
NORMAL 2-Warning 不知道怎么弄 你好,我是柯达的工程师,5800的36935的错误,一般是软件损坏了!
以后又问题可以问我啊,呵呵
我QQ376184604 问题解决了吗??怎么搞好的!!