寻找平衡感 发表于 2009-8-6 20:48:52

球管标定

本帖最后由 无心人 于 2009-8-6 20:50 编辑

3C.7.0GENERATOR CONFIGURATION 发生器配置


* GEN CONFIGURATION *
TUBE SELECTION(球管选择)        AEC SETUP   (AEC设置)   
GENERATOR LIMITS(发生器限制)        AEC CALIBRATION    (AEC标定)
RECEPTOR SETUP(接收器设置)        TUBE CALIBRATION    (球管标定)
I/O CONFIGURATION(i/o配置)        DAP SETUP





EXIT



3C.7.1Tube Selection球管选择

TUBE SELECTION 球管选择功能允许中选要求的管型,允许设置该球管默认限制。

控制台TUBE SELECTION 球管选择菜单的第一个屏幕显示如下。



球管选择屏幕1






* TUBE 1 SELECTION *

A192B0.6/1.2        G2560.6/1.0
A2560.6/1.0        G2920.6/1.2
A2920.6/1.2        G10820.3/1.0
A2720.3/0.6        RAD81.0/2.0





EXIT        >>


球管选择屏幕2




* TUBE 1 SELECTION *

RAD140.6/1.2        RAD740.6/1.5
RAD210.6/1.2        RAD920.6/1.2
RAD560.6/1.2        DX10HS0.6/1.0
RAD600.6/1.2        DX92HS0.6/1.2





<<        >>



CMP 200 X-Ray Generator Service Manual        Ch # 740987-04        Rev. N        Page 3C-19

3CInterfacing and Programming        CPI Canada Inc.


3C.7.1Tube Selection (cont)


球管选择屏幕3



* TUBE 1 SELECTION *

DX93HS0.6/1.5        RO17500.6/1.3
DX101HS0.6/1.3
MX751.0/2.0
MX1000.6/1.25






<<        >>

以下菜单只用在球管型号已经经由电脑运行GenWare下传好时


球管选择屏幕4






TUBE 1 SELECTION *



IF ADDITIONAL TUBES HAVE BEEN DOWNLOADED FROM GenWare
&reg;THEY WILL BE DISPLAYED ON THIS SCREEN
如果附加的管已经从GenWare下载,他们将显示在这屏幕上。






       <<        RETURN



下三个菜单显示球管限制默认值。

球管默认屏幕1
*TUBE 1:        RAD60        0.6/1.2        12&ordm;        REV 1.4*
TUBE SPEED: LOW(球管速度低/高)
MAX SF KW LS:   21.3 (小焦点最大功率)   
MAX LF KW LS:   58.9(大焦点最大功率)        +
MAX KV: 125         (最高千伏)        -







EXIT        >>










球管默认屏幕2
*TUBE 1:        RAD60        0.6/1.2        12&ordm;        REV 1.4*
SF STANDBY: 2.5A(小焦点预热电流)        FIL BOOST:200MS (灯丝增温)
LF STANDBY: 2.5A (大焦点预热电流)        FIL PREHEAT:800MS(灯丝预增温)
SF MAX: 5.2A      (小焦点最大电流)        +
LF MAX: 5.5A      (大焦点最大电流)        -







<<        >>


球管默认屏幕3
*TUBE 1:        RAD60        0.6/1.2        12&ordm;        REV 1.4*
MAX SF MA:        320 (小焦点最大管电流)
ANODE HU WARNING:80%(阳极热报警)
ANODE HU LIMIT:90%   (阳极热限制)        +
-






RETUR


3C.7.2Generator Limits 发生器限制

GENERATOR LIMITS 发生器限制功能允许发生器设置低于输出极限。显示如下。

* GENERATOR LIMITS*
MAX KW:(最大功率)        32      ROTOR BOOST:1800 MS(旋阳启       动)         
MAX MA: (最大管电流)        320
MIN MA:(最小管电流)        10        +
MAX MAS(最大毫安秒)        500                                    -






EXIT





3C.7.3Receptor Setup 接收器设置

The RECEPTOR SETUP function allows each of the image receptors to be programmed as defined in the table following the example menu screens.

MENU 1(菜单1)显示如下





TUBE:        1


* RECEPTORSETUP *

RECEPTOR SYM:
MEMORY: DEF(存储:默认)        +
-






EXIT        >>



MENU 2


* RECEPTOR SETUP *

SF/LF SWITCH:MAN(大小焦转换手/自动)        DEFAULTS
AEC BACKUP:   FIXED(AEC阻止:固定)        AEC CHANNEL:1 ( AEC通道)   
AEC BACKUP MAS:100(AEC阻止毫安秒)        +
AEC BACKUP MS:1000(AEC阻止时间)        -







<<


NOTE:        THE DEFAULTS SELECTION IN MENU 2 IS ONLY AVAILABLE IF MEMORY IN MENU 1 WAS SET
TO DEF.
RECEPTORMENUS3AND4(FOLLOWING)AREONLYACCESSIBLEIFDEFAULTSIS
ENABLED.




















1

        .




MENU 3


* RECEPTOR SETUP DEFAULTS*

TECHNIQUE: AEC        LEFT FIELD: YES(左射野)
FOCUS: SMALL (焦点)        CENTER FIELD: YES(中射野)
FILM SCREEN: 1        RIGHT FIELD: YES(右射野)








<<        >>


MENU 4


* RECEPTOR SETUP DEFAULTS*

KV:        75        DENSITY:        0
MA:        320
MS:        50        +
-







<<





































c.


3C.7.4I/O Configuration(输入输出配置)

The I/O CONFIGURATION function allows programming the states of the exposure for the inputs and the outputs on the room interface board.

The I/O CONFIGURATION menus for the membrane console are shown below.
* INPUTS *
TUBE THERMAL SW.(热保护)_ _ _ _        STANDBY STATE
DOOR ITLK:(门开关)        _ _ _ _









EXIT        >>


* INPUTS *
INTERLOCK 1:(连锁 1)       _ _ _ _        STANDBY STATE
INTERLOCK 2:(连锁 2)        _ _ _ _
BUCKY 1 RDY:(滤线器1准备)        _ _ _ _
BUCKY 2 RDY:(滤线器2准备)        _ _ _ _






<<        ⇑        >>


* OUTPUTS *
BUCKY 1 START:(滤线器1启动)_ _ _ _        STANDBY STATE
BUCKY 2 START:(滤线器2启动)        _ _ _ _
ROOM LIGHT:    (房门灯)        _ _ _ _








<<        ⇑        >>


















3C.7.4I/O Configuration (cont)Definitions of I/O CONFIGURATION menu items.




FUNCTION
(MEMBRANE CONSOLE)
TUBE THERMAL SW


FUNCTION
(GenWare&reg;)

Tube Thermal
Switch


DESCRIPTION



热控开关编程在高压辅助板的J7-1和J7-2

DOOR ITLK        Room Door
Interlock

门联锁装置输入编程在高压辅助版J4-9和J4-10


INTERLOCK 1        Interlock 1        联锁#1输入编程在高压辅助板的J2-3和J2-4

INTERLOCK 2        Interlock 2        联锁#2输入编程在高压辅助板J2-1和J2-2

BUCKY 1 RDY        滤线板1准备        滤线板1输入编程在高压辅助板J2-5和J2-6。

BUCKY 2 RDY        Bucky 2 Ready        Programs the Bucky 2 input at J4-5 and J4-6 on the H.V. auxiliary board.

BKY 1 START        滤线板1起动        Programs the Bucky 1 “start” output at J2-7 on the H.V. auxiliary board.

BKY 2 START        Bucky 2 Start        Programs the Bucky 2 “start” output at J4-1 on the H.V. auxiliary board.

ROOM LIGHT        室内灯        Programs the room light output at J11-3 and J11-4 on the
H.V. auxiliary board.



先前面板定义的输入和输出依下列各项编程
被编程的输入在发生器的状态是活动的或者停止的.停止的输入被忽略。不用的输入将被停止编程.
•        Outputs may be programmed such that the relay connected to the selected output is energized or de- energized during various states of the generator. Unused outputs should normally be programmed to be
de-energized.
•        输入与输出必须编程分离的受体。各受体拥有唯一的编程。






















1



3.


3C.7.5AEC Setup

TAEC SETUP 菜单允许编程各AEC 通道AEC 参数.AEC SETUP 菜单显示如下

MENU 1




* AEC SETUP*

CHANNEL:        1(通道选择)        CHAMBER TYPE:   ION
LEFT FIELD:        YES(左射野)        FILM SCREEN1:YES
CENTER FIELD:        YES(中射野)        FILM SCREEN   2 :YES
RIGHT FIELD:        YES(右射野)        FILM SCREEN3 :YES






EXIT        >>


MENU 2. 在此时不要校准这些值



* AEC SETUP*
R FIELD COMP:        0
C FIELD COMP:        0
L FIELD COMP:        0        +
-



<<




























Page 3C-46        CMP 200 X-Ray Generator Service Manual        Ch # 740987-04        Rev. N



3C.7.7Tube Calibration

球管标定开始以前,用于此装置的球管必须被适当选择,发生器极限应该编程,如前所述。球管自动标定时它是球管推荐条件,如果没有该型号,请参考第六章。

       。
   
球管自动标定步骤如下:

1.        从GENERATOR SETUP菜单选择 GEN CONFIGURATION.

2        选择 TUBE CALIBRATION.

3.        选择焦点。

4.   按下手闸开始标定。

5.    当结束小焦点标定,按RETURN 标定大焦点。
6.   自动标定完毕,按 EXIT 返回到GENERATOR SETUP菜单.
7.   按 EXIT SETUP退出标定模式,返回正常模式。

无心人 发表于 2009-8-6 20:52:26

请你把格式重新排下,其次内容比较多时,可以以附件形式上传上来。还有记得文章的分类。

zhengxinydc 发表于 2011-5-27 11:34:13

顶一个!再顶一个!

houbr 发表于 2011-5-29 21:33:30

本帖最后由 houbr 于 2011-5-29 21:36 编辑

本人这里有该文档的本人注释的原版,需要的同行,请发帖。或qq联系。

mingdr 发表于 2011-6-3 22:19:05

好东西,给我一份,发到我的QQ邮箱里,谢谢,QQ409132669

384659517 发表于 2012-1-31 14:19:26

这是什么设备的资料啊,麻烦说明一下好吗?

longzhen03 发表于 2012-2-2 22:09:21

回楼上,是“CMP 200 X-Ray Generator Service Manual”,CMP200系列X光系统服务手册,这套系统是加拿大的,原版是英文滴,,楼主译了一部分。

lanhaijunjah 发表于 2012-2-5 12:39:02

CMP200发生器调试手册发生器是加拿大CMP公司生产的

涛声依旧_FHKJ 发表于 2018-10-29 23:16:03

学习的看看
页: [1]
查看完整版本: 球管标定