球管标定
本帖最后由 无心人 于 2009-8-6 20:50 编辑3C.7.0GENERATOR CONFIGURATION 发生器配置
* GEN CONFIGURATION *
TUBE SELECTION(球管选择) AEC SETUP (AEC设置)
GENERATOR LIMITS(发生器限制) AEC CALIBRATION (AEC标定)
RECEPTOR SETUP(接收器设置) TUBE CALIBRATION (球管标定)
I/O CONFIGURATION(i/o配置) DAP SETUP
EXIT
3C.7.1Tube Selection球管选择
TUBE SELECTION 球管选择功能允许中选要求的管型,允许设置该球管默认限制。
控制台TUBE SELECTION 球管选择菜单的第一个屏幕显示如下。
球管选择屏幕1
* TUBE 1 SELECTION *
A192B0.6/1.2 G2560.6/1.0
A2560.6/1.0 G2920.6/1.2
A2920.6/1.2 G10820.3/1.0
A2720.3/0.6 RAD81.0/2.0
EXIT >>
球管选择屏幕2
* TUBE 1 SELECTION *
RAD140.6/1.2 RAD740.6/1.5
RAD210.6/1.2 RAD920.6/1.2
RAD560.6/1.2 DX10HS0.6/1.0
RAD600.6/1.2 DX92HS0.6/1.2
<< >>
CMP 200 X-Ray Generator Service Manual Ch # 740987-04 Rev. N Page 3C-19
3CInterfacing and Programming CPI Canada Inc.
3C.7.1Tube Selection (cont)
球管选择屏幕3
* TUBE 1 SELECTION *
DX93HS0.6/1.5 RO17500.6/1.3
DX101HS0.6/1.3
MX751.0/2.0
MX1000.6/1.25
<< >>
以下菜单只用在球管型号已经经由电脑运行GenWare下传好时
球管选择屏幕4
TUBE 1 SELECTION *
IF ADDITIONAL TUBES HAVE BEEN DOWNLOADED FROM GenWare
®THEY WILL BE DISPLAYED ON THIS SCREEN
如果附加的管已经从GenWare下载,他们将显示在这屏幕上。
<< RETURN
下三个菜单显示球管限制默认值。
球管默认屏幕1
*TUBE 1: RAD60 0.6/1.2 12º REV 1.4*
TUBE SPEED: LOW(球管速度低/高)
MAX SF KW LS: 21.3 (小焦点最大功率)
MAX LF KW LS: 58.9(大焦点最大功率) +
MAX KV: 125 (最高千伏) -
EXIT >>
球管默认屏幕2
*TUBE 1: RAD60 0.6/1.2 12º REV 1.4*
SF STANDBY: 2.5A(小焦点预热电流) FIL BOOST:200MS (灯丝增温)
LF STANDBY: 2.5A (大焦点预热电流) FIL PREHEAT:800MS(灯丝预增温)
SF MAX: 5.2A (小焦点最大电流) +
LF MAX: 5.5A (大焦点最大电流) -
<< >>
球管默认屏幕3
*TUBE 1: RAD60 0.6/1.2 12º REV 1.4*
MAX SF MA: 320 (小焦点最大管电流)
ANODE HU WARNING:80%(阳极热报警)
ANODE HU LIMIT:90% (阳极热限制) +
-
RETUR
3C.7.2Generator Limits 发生器限制
GENERATOR LIMITS 发生器限制功能允许发生器设置低于输出极限。显示如下。
* GENERATOR LIMITS*
MAX KW:(最大功率) 32 ROTOR BOOST:1800 MS(旋阳启 动)
MAX MA: (最大管电流) 320
MIN MA:(最小管电流) 10 +
MAX MAS(最大毫安秒) 500 -
EXIT
3C.7.3Receptor Setup 接收器设置
The RECEPTOR SETUP function allows each of the image receptors to be programmed as defined in the table following the example menu screens.
MENU 1(菜单1)显示如下
TUBE: 1
* RECEPTORSETUP *
RECEPTOR SYM:
MEMORY: DEF(存储:默认) +
-
EXIT >>
MENU 2
* RECEPTOR SETUP *
SF/LF SWITCH:MAN(大小焦转换手/自动) DEFAULTS
AEC BACKUP: FIXED(AEC阻止:固定) AEC CHANNEL:1 ( AEC通道)
AEC BACKUP MAS:100(AEC阻止毫安秒) +
AEC BACKUP MS:1000(AEC阻止时间) -
<<
NOTE: THE DEFAULTS SELECTION IN MENU 2 IS ONLY AVAILABLE IF MEMORY IN MENU 1 WAS SET
TO DEF.
RECEPTORMENUS3AND4(FOLLOWING)AREONLYACCESSIBLEIFDEFAULTSIS
ENABLED.
1
.
MENU 3
* RECEPTOR SETUP DEFAULTS*
TECHNIQUE: AEC LEFT FIELD: YES(左射野)
FOCUS: SMALL (焦点) CENTER FIELD: YES(中射野)
FILM SCREEN: 1 RIGHT FIELD: YES(右射野)
<< >>
MENU 4
* RECEPTOR SETUP DEFAULTS*
KV: 75 DENSITY: 0
MA: 320
MS: 50 +
-
<<
c.
3C.7.4I/O Configuration(输入输出配置)
The I/O CONFIGURATION function allows programming the states of the exposure for the inputs and the outputs on the room interface board.
The I/O CONFIGURATION menus for the membrane console are shown below.
* INPUTS *
TUBE THERMAL SW.(热保护)_ _ _ _ STANDBY STATE
DOOR ITLK:(门开关) _ _ _ _
EXIT >>
* INPUTS *
INTERLOCK 1:(连锁 1) _ _ _ _ STANDBY STATE
INTERLOCK 2:(连锁 2) _ _ _ _
BUCKY 1 RDY:(滤线器1准备) _ _ _ _
BUCKY 2 RDY:(滤线器2准备) _ _ _ _
<< ⇑ >>
* OUTPUTS *
BUCKY 1 START:(滤线器1启动)_ _ _ _ STANDBY STATE
BUCKY 2 START:(滤线器2启动) _ _ _ _
ROOM LIGHT: (房门灯) _ _ _ _
<< ⇑ >>
3C.7.4I/O Configuration (cont)Definitions of I/O CONFIGURATION menu items.
FUNCTION
(MEMBRANE CONSOLE)
TUBE THERMAL SW
FUNCTION
(GenWare®)
Tube Thermal
Switch
DESCRIPTION
热控开关编程在高压辅助板的J7-1和J7-2
DOOR ITLK Room Door
Interlock
门联锁装置输入编程在高压辅助版J4-9和J4-10
INTERLOCK 1 Interlock 1 联锁#1输入编程在高压辅助板的J2-3和J2-4
INTERLOCK 2 Interlock 2 联锁#2输入编程在高压辅助板J2-1和J2-2
BUCKY 1 RDY 滤线板1准备 滤线板1输入编程在高压辅助板J2-5和J2-6。
BUCKY 2 RDY Bucky 2 Ready Programs the Bucky 2 input at J4-5 and J4-6 on the H.V. auxiliary board.
BKY 1 START 滤线板1起动 Programs the Bucky 1 “start” output at J2-7 on the H.V. auxiliary board.
BKY 2 START Bucky 2 Start Programs the Bucky 2 “start” output at J4-1 on the H.V. auxiliary board.
ROOM LIGHT 室内灯 Programs the room light output at J11-3 and J11-4 on the
H.V. auxiliary board.
先前面板定义的输入和输出依下列各项编程
被编程的输入在发生器的状态是活动的或者停止的.停止的输入被忽略。不用的输入将被停止编程.
• Outputs may be programmed such that the relay connected to the selected output is energized or de- energized during various states of the generator. Unused outputs should normally be programmed to be
de-energized.
• 输入与输出必须编程分离的受体。各受体拥有唯一的编程。
1
3.
3C.7.5AEC Setup
TAEC SETUP 菜单允许编程各AEC 通道AEC 参数.AEC SETUP 菜单显示如下
MENU 1
* AEC SETUP*
CHANNEL: 1(通道选择) CHAMBER TYPE: ION
LEFT FIELD: YES(左射野) FILM SCREEN1:YES
CENTER FIELD: YES(中射野) FILM SCREEN 2 :YES
RIGHT FIELD: YES(右射野) FILM SCREEN3 :YES
EXIT >>
MENU 2. 在此时不要校准这些值
* AEC SETUP*
R FIELD COMP: 0
C FIELD COMP: 0
L FIELD COMP: 0 +
-
<<
Page 3C-46 CMP 200 X-Ray Generator Service Manual Ch # 740987-04 Rev. N
3C.7.7Tube Calibration
球管标定开始以前,用于此装置的球管必须被适当选择,发生器极限应该编程,如前所述。球管自动标定时它是球管推荐条件,如果没有该型号,请参考第六章。
。
球管自动标定步骤如下:
1. 从GENERATOR SETUP菜单选择 GEN CONFIGURATION.
2 选择 TUBE CALIBRATION.
3. 选择焦点。
4. 按下手闸开始标定。
5. 当结束小焦点标定,按RETURN 标定大焦点。
6. 自动标定完毕,按 EXIT 返回到GENERATOR SETUP菜单.