金宝扳手错误
扳手报错,代码088048怎么处理啊?谢谢 血线探测器调校,一般能过,不行的话,可能是里面的玻璃管脏了,拆开清洗看看。刚工作,希望其他人补充下。 smwj1234v 发表于 2012-3-5 22:45 static/image/common/back.gif
血线探测器调校,一般能过,不行的话,可能是里面的玻璃管脏了,拆开清洗看看。刚工作,希望其他人补充下。
谢谢了,学习了 C COFB 088 048 PRIMING DETECTOR TEST ERROR,
BLOOD IN PRIMING DETECTOR (SECOND TEST), CPU P
Unconditional error, blood pump not available
General description of conditions for occurring:
- CPU P sensed “blood” when the priming detector test intensity is high again
after the low intensity test..
Technical description of conditions for occurring:
- PRI(PI_PrimDetEmptyP2ErrCnt) > 10
Some possible explanations why the error has occurred:
- Priming detector not connected to FM I/O board (P504)
- Software error
- Priming detector is not empty or priming detector dirty
- Priming detector board is broken
- Priming detector transmitter electronics broken (IC55, IC33, Q9, Q13 and
components) on protective driver board.
- Priming detector receiver electronics broken (IC18, V11, V12 and
components) on protective driver board.
C COFB 088 049 PRIMING DETECTOR TEST ERROR,
BLOOD IN PRIMING DETECTOR (SECOND TEST), CPU W 一头雾水,不能整个中文的??
预冲探测器测试错误,预冲探测器 ( 第二次测试) 中有血液,CPU P。 无条件错误, 血泵停止工作。发生的情况一般描述:- 当预冲探测器低强度测试后测试强度再一次升高时,CPU P 感知了 " 血 ";发生的情况技术上描述:- PRI(PI_PrimDetEmptyP2ErrCnt)>10;发生错误的一些可能的解释:- 预冲探测器没有连接FM I/O板(P504)。- 软件错误。- 预冲探测器不是空的或预冲探测器肮脏的。- 预冲探测器板损坏。-在保护驱动板上预冲探测器发射器电子学坏掉 (IC55 , IC33 , Q9 , Q13 和部件)。-在保护驱动板上预冲探测器接收器电子学坏掉 (IC18 , V11 , V12 和部件). 好东西,就是没币了 学习,很好。。。 努力学习。。。。 好东西,学习了。。。。。
页:
[1]
2